Agilent 1200 Preparative Pump User Manual 45
Installing the Pump
3
Flow Connections for a Dual (Binary Gradient) Preparative Pump
Flow Connections for a Dual (Binary Gradient) Preparative Pump
Tools required Wrench 1/4 - 5/16 inch for capillary connections
Parts required # Part number Description
Other modules
1 G1361-68708 Parts from basis kit (see Ta b l e 8 on page 32)
1 G1361-68707 Parts from gradient kit (see Ta b l e 9 on page 33)
Preparations Pumps are installed in the HPLC system
WARNING
When opening capillary or tube fittings solvents may leak out.
The handling of toxic and hazardous solvents and reagents can hold health risks.
Please observe appropriate safety procedures (for example, goggles, safety gloves
and protective clothing) as described in the material handling and safety data sheet
supplied by the solvent vendor, especially when toxic or hazardous solvents are
used.
WARNING
Danger of Explosion
Pumping solvents at high flow rates generally results in electrostatic charging of
the solvents, which may result in the potential ignition of flammable vapors or
solvents.
To avoid electrostatic discharge, never use the pump without grounded metal
solvent bottles.
Do not use metal bottles without appropriate grounding!

Table of Contents

Other manuals for Agilent Technologies 1200 series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 1200 series and is the answer not in the manual?

Agilent Technologies 1200 series Specifications

General IconGeneral
BrandAgilent Technologies
Model1200 series
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals