26 Agilent 1200 Preparative Pump User Manual
2 Site Requirements and Specifications
Site Requirements
Bench Space
The module dimensions and weight (see Table 4 on page 27) allow to place the
module on almost any laboratory bench. It needs an additional 2.5 cm
(1.0 inches) of space on either side and approximately 8 cm (3.1 inches) in the
rear for the circulation of air and electric connections.
If the bench should carry a complete Agilent 1200 Series system, make sure
that the bench is designed to carry the weight of all the modules.
Environment
Your module will work within specifications at ambient temperatures and
relative humidity as described in Table 4 on page 27.
The module should be operated in a horizontal position!
For a complete system including multiple pumps, it is recommended to position the
modules in two or more stacks.
Condensation within the module
Condensation will damage the system electronics.
➔ Do not store, ship or use your module under conditions where temperature
fluctuations could cause condensation within the module.
➔ If your module was shipped in cold weather, leave it in its box and allow it to warm
slowly to room temperature to avoid condensation.