68 Agilent 1200 Preparative Pump User Manual
5 Optimizing Performance
How to Optimize the Compressibility Compensation Setting
Tabl e 11 Solvent Compressibility
Solvent (pure) Compressibility (10-6/bar)
Acetone 126
Acetonitrile 115
Benzene 95
Carbon tetrachloride 110
Chloroform 100
Cyclohexane 118
Ethanol 114
Ethyl acetate 104
Heptane 120
Hexane 150
Isobutanol 100
Isopropanol 100
Methanol 120
1-Propanol 100
To l u e n e 8 7
THF 95
Water 46