Agilent 1200 Preparative Pump User Manual 77
Troubleshooting and Diagnostics
6
Error Messages
Synchronization Lost
During an analysis, the internal synchronization or communication between
one or more of the modules in the system has failed.
The system processors continually monitor the system configuration. If one or
more of the modules is no longer recognized as being connected to the system,
the error message is generated.
Probable cause Suggested actions
1 CAN cable disconnected. • Ensure all the CAN cables are connected
correctly.
• Ensure all CAN cables are installed correctly.
2 Defective CAN cable. Exchange the CAN cable.
3 Defective main board in a different module. Switch off the system. Restart the system, and
determine which module or modules are not
recognized by the system.