EasyManuals Logo

Agilent Technologies 1260 Infinity Quick Start Guide

Agilent Technologies 1260 Infinity
98 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #56 background imageLoading...
Page #56 background image
56 Purification Solution - Developer's Guide
6
Calibration Procedures
Characterizing the delay volumes: Mixing point to UV detector and Column to UV detector
2 Calculate the final Mixing point to UV detector tubing volume as the sum
of:
Mixing point to column tubing volume: step 1
Column to UV detector tubing volume: see “Calculation of delay volume:
Column to UV detector” on page 50
Dual loop autosampler (G2258A), 50 µL loop 0.05
Autosampler to 2/10 valve (0.5 × 600 mm) 0.118
Valve to column (0.17 × 400 mm) 0.009
Total Analytical (mixing point to column)
p
0
/2760 + 1.053
1
For different pump use volumes from tables in Standard analytical system section.
Tabl e 10 UV/MSD-based instrument with UV detector in front of splitter (that is, in
preparative flow) or UV-based preparative instrument
Item Volume, mL
T-connection 0.002
Capillary to mixer (0.6 × 40 mm) 0.011
Mixer (79835-87330) 0.75
Capillary to autosampler (0.6 × 400 mm) 0.113
Dual loop utosampler(G2258A), 5 mL loop 5.13
Capillary to Column (0.5 × 800 mm) 0.156
Total Preparative (mixing point to column): 6.162
Tabl e 9 Analytical flow path of combined instrument:
Item Volume, mL

Table of Contents

Other manuals for Agilent Technologies 1260 Infinity

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 1260 Infinity and is the answer not in the manual?

Agilent Technologies 1260 Infinity Specifications

General IconGeneral
BrandAgilent Technologies
Model1260 Infinity
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals