46 1260 Infinity Micro Degasser Service Manual
6 Troubleshooting and Diagnostics
Status Indicators
defect that does not allow the vacuum to be kept below the error
threshold (180 Torr).
CAUTION
Error LED is on
This indicates either an internal leak in the vacuum system or an electronic failure.
In case of an internal leak it is possible that solvent may enter the vacuum chamber
and solvent may leak into the waste drain.
To prevent any damage, switch off the micro vacuum degasser and remove the
solvent bottles from the solvent cabinet to stop any gravity-caused flow of solvent
into the vacuum chamber.