1260 Infinity Micro Degasser Service Manual 61
Maintenance
7
Introduction to Maintenance and Repair
3 Press down the clip of the fuse holder and pull out of the power socket.
Figure 12 Fuse Holder and Clip
4 Remove the fuses from the fuse holders.
5 Ensure the fuse wires inside the fuses are not broken. If a test meter is
available, check the resistance of each fuse. A good fuse shows a low
resistance (approximately 0 Ohm).
6 If a fuse is defective (wire broken or high resistance), insert a new fuse.
7 Reinsert the fuse holders and the power cable.
8 Switch on the power switch.