12 Agilent 2200 TapeStation System - User Manual
2 Specifications
Technical Specifications
Technical Specifications
2200 TapeStation instrument
Input voltage: 12 V DC
Power consumption: 40 W (VA)
Current: 3 A
Interface: USB cable (PC comms.)
Instrument Housing: UL94/VO rated flame retardant cast polyurethane
Dimensions: 400x 310x 310 mm
Weight: 12.5 kg
Power supply
Input voltage: 100 – 240 V AC
Input frequency: 50 – 60 Hz
Phase: 1
Current: 0.45 – 1.1 A
Environmental conditions
Optimal operating
temperature
20 °C (68 °F).
Instrument operating
temperature
1
1
Instrument operating temperature may be higher than ambient lab temperature, especially after
prolonged use.
12 – 37 °C (54 – 99 °F) for D1000 ScreenTape assay
17 – 37 °C (63 – 99 °F) for High Sensitivity D1000 ScreenTape assay
15 – 30 °C (59 – 86 °F) for Genomic DNA, D5000 and High Sensitivity
D5000 ScreenTape assays
14 – 30 °C (57 – 86 °F) for RNA and High Sensitivity RNA ScreenTape
assays
10 – 33 °C (50 – 91 °F) for P200 ScreenTape assay