Agilent 2200 TapeStation System - User Manual 55
Using the 2200 TapeStation System
4
DNA Sample Preparation
Sample Preparation D1000 ScreenTape Assay
Prepare TapeStation System D1000
1 Launch the 2200 TapeStation Controller Software.
2 Load D1000 ScreenTape device and loading tips into the 2200
TapeStation instrument.
Sample Preparation D1000 ScreenTape Assay
1 Allow reagents to equilibrate at room temperature for 30 min
2 Vor tex mix before use
3 If running ladder, prepare by mixing 3 µL D1000 Sample Buffer ( )
with 1 µL D1000 Ladder ( )
4 Prepare sample by mixing 3 µL D1000 Sample Buffer ( ) with 1 µL
DNA sample
5 Spin down, then vortex using IKA vortexer and adaptor at 2000 rpm
for 1 min.
6 Spin down to position the sample at the bottom of the tube.
Parts required p/n Description
5067-5582 D1000 ScreenTape
Parts required p/n Description
5067-5583 D1000 Reagents
&¥A9C6
dgAVYYZg
(¥A9&%%%
HVbeaZ
7j[[Zg
KdgiZm&b^c
He^cYdlc
9&%%%/

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 2200 and is the answer not in the manual?

Agilent Technologies 2200 Specifications

General IconGeneral
BrandAgilent Technologies
Model2200
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals