14 Agilent 2200 TapeStation System - User Manual
2 Specifications
ScreenTape Specifications
ScreenTape Specifications
Specification (D1000 ScreenTape Assay)
Analytical Specifications D1000 ScreenTape assay and reagents
Sizing Range 35 – 1000 bp
Typical Resolution 35 – 300 bp: 15 %
300 – 1000 bp: 10 %
Sensitivity
1
1
signal-to-noise >3 (single peak)
0.1 ng/µL
Sizing Precision 5 % CV
Sizing Accuracy
2
2
Sizing Accuracy for electronic ladder: ±20%
±10 %
Quantitative Precision 0.1 – 1 ng/µL: 15 % CV
1 – 50 ng/µL: 10 % CV
Quantitative Accuracy
3
3
Measured against 2100 Bioanalyzer system
±20 %
Quantitative Range 0.1 – 50 ng/µL
Maximum sample buffer strength 20 mM KCl,
60 mM Phosphate Buffer,
60 mM Guanidine-HCl,
240 mM NaCl,
60 mM Acetate
Physical Specifications
Analysis Time 16 samples: <20 min
96 samples: ≈100 min
Samples per consumable 16
Sample volume required 1 µL
Kit stability 4 months
Kit size 112 samples