56 Agilent 2200 TapeStation System - User Manual
4 Using the 2200 TapeStation System
DNA Sample Preparation
Sample Preparation High Sensitivity D1000 ScreenTape Assay
Prepare TapeStation System HSD1000
1 Launch the 2200 TapeStation Controller Software.
2 Load High Sensitivity D1000 ScreenTape device and loading tips into
the 2200 TapeStation instrument.
Sample Preparation High Sensitivity D1000 ScreenTape Assay
1 Allow reagents to equilibrate at room temperature for 30 min
2 Vor tex mix before use
3 If running ladder, prepare by mixing 2 µL High Sensitivity D1000
Sample Buffer ( ) with 2 µL High Sensitivity D1000 Ladder ( )
4 Prepare sample by mixing 2 µL High Sensitivity D1000 Sample Buffer
() with 2µL DNA sample
5 Spin down, then vortex using IKA vortexer and adaptor at 2000 rpm
for 1 min.
6 Spin down to position the sample at the bottom of the tube.
Parts required p/n Description
5067-5584 High Sensitivity D1000 ScreenTape
Parts required p/n Description
5067-5585 High Sensitivity D1000 Reagents
KdgiZm&b^c
He^cYdlc
'¥A9C6
dgAVYYZg
'¥A9&%%%
HVbeaZ
7j[[Zg
=^\]HZch^i^k^in
9&%%%/

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 2200 and is the answer not in the manual?

Agilent Technologies 2200 Specifications

General IconGeneral
BrandAgilent Technologies
Model2200
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals