Agilent 2200 TapeStation System - User Manual 21
Specifications
2
ScreenTape Specifications
Specification (P200 ScreenTape Assay)
Analytical Specification P200 ScreenTape assay and reagents
Sizing range 10 – 200 kDa
Resolution
1
1
determined using P200 Ladder as sample
15 %
Typical Sizing Accuracy ±10 % (CAII, Lysozyme, beta lactoglobulin)
Sizing Precision 3 % CV
Quantitative Range/precision 100 – 1000 ng/µL for IgG; 15 % CV
Qualitative Range 5 – 5000 ng/µL BSA, Lysozyme; 12.5 – 5000 ng/µL IgG
Sensitivity
2
2
signal :noise ratio > 3
5 ng/µL Lysozyme; 12.5 ng/µL IgG
Physical Specification
Sample volume needed 2 µL
Analysis Time 16 samples: <15 min
Samples/consumable 16
Kit Size 112 Samples
Kit Stability 4 months