4349B
Figure
9-3.
Ammeter
Input
Resistance
T
est
Setup
Using
DC
V
oltage
Calibrator
14.
Set
the
RC
Box
resistor
to
10
4
.
15.
Set
the
DC
power
supply
(voltage
calibrator)
to
1
V
.
16.
Record
the
multimeter
reading
on
the
calculation
sheet
on
the
L
V
oltage
line
.
17.
Disconnect
the
Dual
Banana-BNC(m)
cable
from
A
of
Figure
9-2
,
and
connect
it
to
B
.
If
the
voltage
calibrator
is
used,
skip
this
step
.
18.
Record
the
multimeter
reading
(voltage
calibrator
setting)
on
the calculation
sheet on
the
H
V
oltage
line
.
19.
Calculate
the
test
result
according
to
the
calculation
sheet, and
record the
result
into
the
performance
test
record.
20.
Remove the
triaxial cable
from the
4349B
's
channel
1
input
and
connect
it
to
channel
2
input.
21.
Test
channel 2
using the
steps 13
to
19.
22.
T
est
channel
3
and
channel
4
using
the
same
procedure.
If
the
4349B
is
equipped
with
Option 001 (Two T
est
Channels), skip this step
.
Maintenance 9-5

Table of Contents

Other manuals for Agilent Technologies 4349B

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 4349B and is the answer not in the manual?

Agilent Technologies 4349B Specifications

General IconGeneral
BrandAgilent Technologies
Model4349B
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals