4349B
T
o
Set
the
Contact
Check
The
Cont Check
annunciator(
9
) tells
the
current
contact
check
status
.
T
o
set
or
abort
the
contact
check function:
Note
Before
performing
the
contact check,
perform the
OPEN correction
to measure
the
reference
value
of the
contact check.
Note
The
contact
check
feature
of
the 4349B
discriminates Contact/NO
Contact
based
on the
dierence in
capacitance between
when the
DUT
is
connected
and
when it
is not
connected (Open).
Therefore,
if
the
change
of
capacitance
between
those
two
dierent states
is not
enough, the
discrimination
may
fail.
T
o
enable
or
disable
the
contact
check
function:
1.
Press
.
2.
Select
ON/OFF
using
or and
press
to
select.
3.
Select
On
or
Off
using
or ,
and
press
to
exit
to
the
previous
display
.
4.
Select
Exit
using
or ,
and press
to
exit.
5.
The
Cont
Chk
annunciator(
9
)
turns
ON
if
the
contact
check
function
is enabled.
In
the
case
of
the
contact
check
function
is
enabled,
N.C.
(No-Contact)
will be
displayed
when
the
contacts
between
dut
and
xture
are
open.
See
\Contact
Check
Eective Range
"
in
Chapter 8 for the contact check eective range
.
To perform the contact check eectively
, note the following:
Connect the coaxial cable as close to the DUT
as possible
.
If the judgment result is Contact when the DUT is
not connected due to variation in residual
capacitance of the xture or cable used for the automated test
equipment, set an Oset
value for the Limit value to provide a stricter discrimination criterion.
For the principle of operation of the contact check, refer to Chapter 3 \Principle of Operation."
Contact check compares a measured capacitance data of dut with a limit data. Interpretation
of the limit data and the measured capacitance data of the dut are shown below.
2-8 Operating the 4349B