4349B
10.
Set the
measurement parameter
to current
(I).
a.
Press
.
b. Select
I
using
or and press .
11.
Press
the
IN5
switch
on
the
Handler
Interface
T
ester
,
and conrm
that the
4349B is
triggered.
Handler
Interface
Output
T
est.
12.
Start
the
handler
interface
output
test
with
the
following
procedure
.
a.
Press
to
display
the
conguration
setting
menu.
b.
Select
Svc
using
or and
press
.
c.
Select
HNDL
using
or and
press
.
13.
Set
the
LED
PW
switch
on
the
Handler
Interface
T
ester
to
ON
.
Caution
Don't
set
the
LED
PW
switch
to
ON
,
except when
doing the
handler
interface
output
test.
The
LEDs
may
cause
the
4349B
to
shut-down
during
normal
operation.
14.
Conrm
that
the
LEDs
on
the
Handler
Interface
T
ester
light
in
the
order
shown
in
Figure
9-11
,
in
accordance
with
the
4349B
display
.
Figure 9-11. Handler interface Output Order
15. Set the
LED PW
switch on the Handler Interface Tester to OFF.
16. Press twice to exit from the test mode.
9-24 Maintenance

Table of Contents

Other manuals for Agilent Technologies 4349B

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 4349B and is the answer not in the manual?

Agilent Technologies 4349B Specifications

General IconGeneral
BrandAgilent Technologies
Model4349B
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals