4349B
External V
oltage Source
The
4349B
has
no
internal test
voltage source
.Y
ou must
supply a
suitable external
voltage
source
in
order
to
make resistance
measurements.
In
a
typical
case
, measuring
the insulation
resistance of
capacitors whose
insulation resistance
value
is
expressed

1
G
1

F within

10%
measurement accuracy
,
the
external
voltage
source
must
meet
the
following
specications:
Ripple
:

1
mV
rms
(

0.7
mV
rms
preferable)
@
50/60
Hz
Wide
Band
Noise
:

5

V
rms
/
p
H
z
(

2

V
rms
/
p
H
z
preferable)
@
50
Hz
Switching
Noise
:

50
mV
rms
T
est Leads
Use
the
16117E
T
est
Lead
to
connect
the
4349B
and the
DUT
.
The
16117E
contains
one
cable
(1
m,
with
triaxial
connectors
on
the
both
ends)
and
one
connector (Triaxial
F
emale).
So
,
four
16117E
Test
Leads are
required
when
you
want
to
use
four
channels
.
Specications
Cable
Length
:
1
m
(triaxial
connector
to triaxial
connector)
Operating
T
emperature
:
0
to
45

C
Operating
Humidity
:

70%
RH
(@
40

C)
Non-operating
T
emperature :
0
40
to
70

C
Non-operating
Humidity
:

95%
RH
(@
40

C)
Figure 2-1. 16117E T
est Lead
2-2 Operating the 4349B

Table of Contents

Other manuals for Agilent Technologies 4349B

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 4349B and is the answer not in the manual?

Agilent Technologies 4349B Specifications

General IconGeneral
BrandAgilent Technologies
Model4349B
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals