4349B
T
o
Set
the
A
veraging
rate
The
following
commands
set
the
averaging
rate:
:SENS:AVER
:SENS:AVER:COUN
F
or
example
,
to
set
the
averaging
rate
to
4,
OUTPUT
717;":SENS:AVER:COUN
4"
To
Set the
Trigger Delay
time
The following
command sets
the trigger
delay time:
:TRIG:DEL
For
example,
to
set
the
trigger
delay
time
to
10
ms
,
OUTPUT 717;":TRIG:DEL
1E-2"
T
o
P
erform
the
Contact
Check
The following
command
performs
the
contact
check
function:
:SENS:RES:CONT:VER
F
or
example
,
OUTPUT
717;":SENS:RES:CONT:VER
ON"
T
o
Set
the
Beeper
Mode
The
following
commands
set
the
beeper
mode:
:SYST:BEEP
:SYST:BEEP:STAT
:CALC{1|2|3|4}:LIM:BEEP
:CALC{1|2|3|4}:LIM:BEEP:COND
F
or
example
,
to
set
the
beeper
mode
to
emit
a
beep
when
comparison result
is P
ASS,
OUTPUT
717;":CALC1:LIM:BEEP:COND
PASS"
T
o
Lock
Out
the Front
P
anel
K
eys
The following command locks out the front panel
keys:
:SYST:KLOC
For example
,
OUTPUT 717;":SYST:KLOC ON"
Remote Operation 4-7

Table of Contents

Other manuals for Agilent Technologies 4349B

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 4349B and is the answer not in the manual?

Agilent Technologies 4349B Specifications

General IconGeneral
BrandAgilent Technologies
Model4349B
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals