4349B
Handler
Interface
Functional
T
est
The
4349B
's
handler
interface
function
is
tested
using
the
built-in selftest
and the
handler
interface
tester
.
T
est
Equipment
Description
Recommended Model
Handler
Interface T
ester 04339-65007
Procedure
Initial
Setup.
1.
Turn the
4349B OFF
.
2.
Set the
LED
PW
switch on
the
Handler
Interface
T
ester
to
OFF
,
and
set
the
IN1,
IN2,
IN3,
and
IN4 switches
to
0
.
3.
Set up
the
equipment
as
shown
in
Figure
9-10
Figure
9-10.
Handler
Interface Functional
Test
Setup
4.
Turn the
4349B ON.
5.
Reset
the
4349B
using
the
following
procedure
.
a.
Press
to
display the
system
reset
settings
.
b. Select
Yes
using or and press .
Key Lock Function T
est.
6. Set the
IN4
switch on the Handler Interface T
ester to
1
.
7. Conrm that the all keys on the 4349B's front panel are locked out.
8. Set the
IN4
switch on the Handler Interface
Tester to
0
.
External Trigger Function Test.
9. Press
to set the trigger mode to External.
Maintenance 9-23

Table of Contents

Other manuals for Agilent Technologies 4349B

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 4349B and is the answer not in the manual?

Agilent Technologies 4349B Specifications

General IconGeneral
BrandAgilent Technologies
Model4349B
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals