4349B
7.
Press
and
select
ON/OFF
,
ON
to turn
the contact
check
function
ON.
8. Conrm
that the
4349B displays
N.C.
(No
Contact).
9. Connect
the
BNC(m)-BNC(m)
adapter
to
the
MONITOR
terminal
of
the
SMU
T
est A
dapter.
10. Conrm
that
the
N.C.
disappears
and
the
4349B
displays
the
measurement
value
.
11.
Press
to
change
channel,
and
test
the
other
channels
using
the
steps
2
to
10.