4349B
Connect capacitors
(C
b
) with
a value
of about
500 pF
(100 to
1000
pF)
between
the
test
voltage side
terminal of
the device
under test
and the
shield
of
the
triaxial
test
cable
to
bypass the
contact check
signal. (When
Rs
is
greater
than
1
k.)
For
the principle
of operation
of
the
contact
check,
refer
to
Chapter
3
\Principle
of
Operation."
Insert a
resistor between
the
test
voltage
source
and
device
under
test.
This
resistor
is
for
limitting the
current and
reducing
the
voltage
source
noise
eect.
The
minimum
value
for
current
limit
is
calculated
as
follows:
Rs
=
T
est
V
oltage
/
Current
Limit
(F
or
details
about
reducing
the
voltage
source
noise
eect,
refer
to
\How
to
Reduce
the
Measurement
Error"
in
Chapter
6.)
Caution
Do
not
apply
a
dc
voltage
exceeding
6
50 V
or
6
50 mA
to
the
INPUT
terminals
,
or
the
4349B
's
input
circuits
will be
destroyed. Connect
a current
limitting
resistor
between
the
test voltage
source and
device under
test
to
avoid
the
over
current
when
a shorted
device is
connected.
The
series
resistance
value
should
be:
Rs
>
(
Test
Voltage[V]
/
0.05[A]
)
W
arning
The
measurement
system
using
the
4349B
and
the
external
voltage
source
may
output
dangerous
voltage
.
T
o
prevent
an
electrical
shock,
observe
the
following
safety
precautions
.
Agilent
T
echnologies
recommends
you
to
attach
the
safety
interlock
function
to
the
measurement
system
so
that
the
operators
can
not
touch
the
parts
where
dangerous
voltage
may
appear
.
If
not
possible
,
warn
the
operators
and
the
workers
around
the
measurement
system
not
to
touch
the
parts
.
Use
the
parts
with
the
safety
function
to
prevent an
electrical shock
caused
by
the
insulation
breakdown,
even
if
dangerous
voltage does
not
appear
on
the
parts
directly
.
Turn
OFF
the
output
of
the
external
voltage
source and
discharge
the
remanent
charge
of
the
DUT
thoroughly
,
before touching
the
measurement system
or the
DUT where
the dangerous
voltage
may
appear.
Use
the
power
cord
shipped
with
the
4349B
and
provide
an
uninterruptible safety earth ground to the instrument.
Do not apply a dc voltage exceeding
6
50 V or
6
50 mA to the INPUT
terminals.
Agilent T
echnologies warrants the 4349B
and its accessaries provided by
Agilent T
echnologies. Agilent T
echnologies
does not warrant any accidents
caused by trouble in an external voltage source
or the measurement
system.
2-4 Operating the 4349B

Table of Contents

Other manuals for Agilent Technologies 4349B

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 4349B and is the answer not in the manual?

Agilent Technologies 4349B Specifications

General IconGeneral
BrandAgilent Technologies
Model4349B
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals