Chapter 3: Testing Performance
To test the dc calibrator
3–4
To test the dc calibrator
The Aux Out BNC on the back panel is used for self-calibration and probe calibration. Though
calibrator accuracy is not specified in the performance specifications, it must be within limits in
order to provide accurate self-calibration.
Test Limits: -2.4 v to +2.4 v, Accuracy ±0.2% of delta voltage output
Equipment Required
Procedure
1 Connect the multimeter to the back panel Aux Out BNC.
Use the BNC cable and the BNC to banana plug adapter.
2 Enable the graphical interface.
Use the mouse to click on the button in the upper right-hand corner of the display. See Figure 3-1.
Figure 3-1
Enabling the graphical interface
3 Press Default Setup to set the oscilloscope to default conditions.
4 Select Calibration from the Utilities menu.
5 Select DC from the Aux Output drop-down list box.
See Figure 3-2.
Equipment Critical Specifications Recommended Model/Part
Digital Multimeter 0.1 mV resolution, better than 0.1% accuracy Agilent 34401A
Cable BNC Agilent 10503A
Adapter BNC (f) to banana (m) Agilent 1251-2277
Click the mouse with the
pointer on the right-hand
button to enable the
graphical interface

Table of Contents

Other manuals for Agilent Technologies 54830 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 54830 Series and is the answer not in the manual?

Agilent Technologies 54830 Series Specifications

General IconGeneral
BrandAgilent Technologies
Model54830 Series
CategoryTest Equipment
LanguageEnglish

Related product manuals