Chapter 5: Troubleshooting
Primary Trouble Isolation
5–9
No D
isplay Trouble Isolation Flowchart
No Display Trouble
Isolation
Remove cabinet and
install fan guard
Check fan
connections and
power-on
Fans running?
Power LED lit?
Connect external
monitor, cycle power,
and check power up
sequence
Power-up
displayed on ext.
monitor
Replace SVGA card
Go to Primary
Troubl
e Isolation
Go to Power Supply
Trouble Isolation
Go to To check the
backlight Inverter
voltages
Video to flat-
panel display
OK?
Yes
No
Yes
No
Replace
Motherboard
Assembly
Yes
Yes
A
B
C
D
E
No
54830F02