EasyManua.ls Logo

Agilent Technologies 7000 Triple Quadrupole GC/MS

Agilent Technologies 7000 Triple Quadrupole GC/MS
214 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
General Troubleshooting 2
Troubleshooting and Maintenance Manual 31
Peaks have split tops
Bad injection technique
Injection is too large
Baseline is rising
Column bleed
Other contamination
Baseline is high
Column bleed
Other contamination
Electron multiplier voltage is too high
Baseline is falling
A falling baseline indicates contamination is being swept away. Wait until the
baseline reaches an acceptable level. Common causes include:
Residual air and water from a recent venting
Column bleed
Septum bleed
Splitless injection time too long (inlet is not properly swept, resulting in
excess solvent on the column and slow solvent decay)

Table of Contents

Related product manuals