32 Troubleshooting and Maintenance Manual
2 General Troubleshooting
Baseline wanders
Insufficient carrier gas supply pressure*
Malfunctioning flow or pressure regulator*
Intermittent leak in the GC inlet*
* This could cause a fault condition in the GC that would prevent the GC
from operating.
Retention times for all peaks drift – shorter
Column has been shortened
Initial oven temperature was increased
Column is getting old
Retention times for all peaks drift – longer
Column flow has been reduced
Initial oven temperature was decreased
Active sites in the sample path
Leaks in the GC inlet*
* This could cause a fault condition in the GC that would prevent the GC
from operating.
Poor sensitivity
Incorrect tuning, or tune file that does not match the type of analysis
Repeller voltage is too low
Incorrect temperatures (oven, GC/MS interface, ion source, or mass filter)
Incorrect sample concentration
Leaking GC inlet*
Dirty GC inlet
Incorrect split ratio
Purge-off time in splitless mode is too short

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 7000 Triple Quadrupole GC/MS and is the answer not in the manual?

Agilent Technologies 7000 Triple Quadrupole GC/MS Specifications

General IconGeneral
BrandAgilent Technologies
Model7000 Triple Quadrupole GC/MS
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals