General Troubleshooting 2
Troubleshooting and Maintenance Manual 33
Excessive pressure in the front or rear chamber
Dirty ion source
Air leaks between chambers
Poor filament operation
Detector (HED electron multiplier) is not working correctly
Incorrect mass filter polarity
Collision cell voltage
* This could cause a fault condition in the GC that would prevent the GC
from operating.
Poor repeatability
Dirty syringe needle
Dirty GC inlet
Leaking GC inlet*
Injection is too large
Loose column connections
Variations in pressure, column flow, and temperature
Dirty ion source
Loose connections in the analyzer
Ground loops
* This could cause a fault condition in the GC that would prevent the GC
from operating.