EasyManua.ls Logo

Agilent Technologies 7000 Triple Quadrupole GC/MS

Agilent Technologies 7000 Triple Quadrupole GC/MS
214 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
46 Troubleshooting and Maintenance Manual
2 General Troubleshooting
There is no emission current
Check tune file to be certain that emission current is not = 0
Filament is not connected properly; try the other filament
Filament has failed; try the other filament
MS electronics are not working correctly
There is not enough signal to begin tune
Corrupted tune file
Poor mass axis calibration
Width gain or offset is too high
Calibration vial(s) empty or almost empty
Excessive pressure in the analyzer chamber
Air leak
Electron multiplier voltage is too low
Signal cable is not connected
Electrical leads to the detector are not connected correctly
Electrical leads to the ion source are not connected correctly
Filament shorted to the source body
Collision cell gas flow
Collision cell voltages

Table of Contents

Related product manuals