EasyManuals Logo

Agilent Technologies 7000 Triple Quadrupole GC/MS User Manual

Agilent Technologies 7000 Triple Quadrupole GC/MS
214 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #83 background imageLoading...
Page #83 background image
General Maintenance 4
Troubleshooting and Maintenance Manual 83
To move or store the MS
Materials needed
Ferrule, blank (5181-3308)
Interface column nut (05988-20066)
Wrench, open-end, 1/4-inch × 5/16-inch (8710-0510)
Procedure
WARNING
Make sure the GC/MS interface and the analyzer zones are cool (below 100 °C)
before you vent the MS. A temperature of 100 °C is hot enough to burn skin; always
wear cloth gloves when handling analyzer parts.
WARNING
If you are using hydrogen as a carrier gas, the carrier gas flow must be off before
turning off the MS power. If the foreline pump is off, hydrogen will accumulate in the
MS and an explosion may occur. Read “Hydrogen Safety” before operating the MS
with hydrogen carrier gas.
WARNING
When the MS is vented, do not put the MassHunter Workstation software into
Instrument Control view. Doing so will turn on the interface heater.
CAUTION
Be sure the GC oven and the GC/MS interface are cool before turning off the carrier
gas flow.
CAUTION
Never vent the MS by allowing air in through either end of the foreline hose. Use the
vent valve or remove the column nut and column.
Do not vent while the turbo pump is still spinning at more than 50%.
Do not exceed the maximum recommended total gas flow. See Ta bl e 2 .

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 7000 Triple Quadrupole GC/MS and is the answer not in the manual?

Agilent Technologies 7000 Triple Quadrupole GC/MS Specifications

General IconGeneral
BrandAgilent Technologies
Model7000 Triple Quadrupole GC/MS
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals