EasyManuals Logo

Agilent Technologies 7000 Triple Quadrupole GC/MS User Manual

Agilent Technologies 7000 Triple Quadrupole GC/MS
214 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #84 background imageLoading...
Page #84 background image
84 Troubleshooting and Maintenance Manual
4 General Maintenance
1 Vent the MS. See the Agilent 7000 Series Triple Quad GC/MS Operation
Manual.
2 Remove the column.
3 Disconnect the collision gas supply tubing.
4 Install a plug on the end of the collision gas tubing that leads into the
analyzer chamber.
5 Move the MS away from the GC.
6 Unplug the GC/MS interface heater cable from the GC.
7 Remove the front analyzer window and open the left side analyzer panel.
See To remove the MS covers on page 80.
8 Finger-tighten the side plate thumbscrews for both analyzers. (Figure 5).
9 Switch the MS on to establish a rough vacuum.
10 When you hear the hissing sound of the pumpdown, close the vent valve.
Continue the pumpdown for 2 to 3 minutes.
11 Switch the MS off.
12 Close the analyzer cover and replace the front analyzer window.
13 Disconnect the LAN, remote, and power cable.
CAUTION
Do not overtighten the side plate thumbscrews. Overtightening will strip the threads in
the analyzer chamber. It will also warp the side plate and cause leaks.
CAUTION
Always wear clean gloves while handling any parts that go inside the analyzer
chambers.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 7000 Triple Quadrupole GC/MS and is the answer not in the manual?

Agilent Technologies 7000 Triple Quadrupole GC/MS Specifications

General IconGeneral
BrandAgilent Technologies
Model7000 Triple Quadrupole GC/MS
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals