EasyManua.ls Logo

Agilent Technologies 7000 Triple Quadrupole GC/MS

Agilent Technologies 7000 Triple Quadrupole GC/MS
214 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
92 Troubleshooting and Maintenance Manual
4 General Maintenance
Reinstallation
Materials needed
Calibration valve (G3170-60204)
O-ring for calibration valve (0905-1217)
PFTBA (05971-60571)
Screwdriver, Torx T-15 (8710-1622)
1 Remove the old valve assembly. See “To remove/reinstall the calibration
and vent valve assemblyon page 91 and Figure 7.
2 Be sure the valve O-ring is in place. If it is worn or damaged, replace it.
3 Install the calibration and vent valve assembly and tighten the screws that
hold it in place.
4 Reconnect the calibration valve cable to the connector.
5 Remove the vial from the new calibration valve. See “To refill/purge the EI
calibration vial” on page 89. The valve is supplied with a vial already
installed.
6 Fill and reinstall the calibration vial. See “To refill/purge the EI
calibration vial” on page 89.
7 Pump down the MS. See the Agilent 7000 Series Triple Quad GC/MS
Operation Manual.
CAUTION
Do not turn on any GC heated zones until the carrier gas flow is on. Heating a column
with no carrier gas flow will damage the column.

Table of Contents

Related product manuals