Chapter 2 2-111
System Verification and Performance Tests
Agilent 8753ET System Verification and Performance Tests
NOTE The analyzer displays the PC annotation, indicating the power meter
calibration is done and the error correction is active.
Output Power Linearity Measurement
1. Set up the equipment as shown in Figure 2-62.
Figure 2-62 Output Power Linearity Test
2. Press .
3. Press .
4. Press .
5. Press
.
6. Press (or the next power setting from the performance test record).
7. Press .
8. Press
.
9. Read the value for marker 1 from the display and enter it in the performance test
record in the “Marker 1 (Maximum)” column. Enter the value of marker 2 in the
“Marker 2 (Minimum)” column.
10.Press
(or the next power setting from the performance test record) .
11.Repeat steps 6 through 10 for all power settings listed on the performance test record in
the 300 kHz to 3 GHz range.
Meas
Display
Marker
300 k/m
−15 x1
∆
∆
Power −15 x1
Marker
Marker Search
Marker
Marker
−13 x1