CALibration Subsystem
CALibration Subsystem
The CALibration subsystem is used to calibrate the instrument.
Commands and Queries
CALibration[:ALL]?
CALibration:COUNt?
CALibration:SECure:CODE <new_code>
CALibration:SECure:STATe {OFF | ON} [,<code>]
CALibration:SECure:STATe?
CALibration:SETup <step>
CALibration:SETup?
CALibration:STORe
CALibration:STRing "<string>"
CALibration:STRing?
CALibration:VALue
CALibration:VALue?
CALibration[:ALL]?
Performs a calibration using the calibration value (CALibration:VALue). The instrument must be unlocked (CAL-
ibration_SECure_STATe OFF,<code>) to calibrate.
Parameter Typical Return
(none) +0 (pass) or +1 (fail)
Calibrate using the current value:
CAL?
l CALibration:SETup should always precede the CALibration? query.
l Increments the instrument’s calibration count (CALibration:COUNt?).
l Modifies the volatile version of the calibration constants. Use CALibration:STORe to save these constants in non-
volatile memory at end of calibration.
CALibration:COUNt?
Returns the number of calibrations performed. Read and record the initial count when you receive your instrument
from the factory.
Parameter Typical Return
(none) +117