EasyManuals Logo
Home>Agilent Technologies>Inverter>Agilent 33500 Series

Agilent Technologies Agilent 33500 Series User Manual

Agilent Technologies Agilent 33500 Series
429 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #386 background imageLoading...
Page #386 background image
General Calibration/Adjustment Procedure
General Calibration/Adjustment Procedure
Recommended method for a complete instrument calibration:
1. Read Test Considerations.
2. Perform the verification tests to characterize the instrument (incoming data).
3. Press [System, then Instr Setup, then Calibrate. If the instrument is secured from calibration, unsecure it.
4. Enter the Setup Number for the procedure being performed. The default setup number is "1" and, from the front
panel, the number will increment as the procedures are performed.
5. Select BEGIN.
6. For setups that require an input, adjust the value shown in the display to the measured value and select ENTER
VALUE.
7. The setup automatically advances to the next required value.
To cancel the adjustment procedure, select CANCEL STEP. The display will return to the setup number entry.
8. When finished, select END CAL.
9. (Optional) Set a new calibration message using the remote interface. The message (up to 40 characters) is stored
with the calibration coefficients.
10. Secure the instrument against calibration.
11. Note the new security code and calibration count in the instrument’s maintenance records.
Agilent 33500 Series Operating and Service Guide 385

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies Agilent 33500 Series and is the answer not in the manual?

Agilent Technologies Agilent 33500 Series Specifications

General IconGeneral
BrandAgilent Technologies
ModelAgilent 33500 Series
CategoryInverter
LanguageEnglish

Related product manuals