EasyManua.ls Logo

Agilent Technologies Agilent 33500 Series

Agilent Technologies Agilent 33500 Series
429 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Output Impedance Adjustment (Channel 2)
Output Impedance Adjustment (Channel 2)
The instrument stores calibration constants for the channels' output impedance. These constants are generated with
and without the post-amplifier attenuator.
1. Set the DMM to measure offset-compensated, four-wire Ohms. Set the DMM to use 100 NPLC integration. Con-
nect the Ohms Source and Ohms Sense DMM inputs to the channel output as shown below.
2. Use the DMM to make a 4-wire resistance measurement at the front panel output connector for each setup in the
following table. The expected measured value is approximately 50 Ω.
Setup
55* -24 dB post-attenuator range
56* 0 dB range
* Constants are stored after completing this setup.
3. Using the numeric keypad or knob, adjust the displayed impedance at each setup to match the measured imped-
ance. Select ENTER VALUE.
4. There are no specific operational verification tests for output impedance. Continue with the next adjustment pro-
cedure in this section.
400 Agilent 33500 Series Operating and Service Guide

Table of Contents

Related product manuals