EasyManuals Logo
Home>Agilent Technologies>Inverter>Agilent 33500 Series

Agilent Technologies Agilent 33500 Series User Manual

Agilent Technologies Agilent 33500 Series
429 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #57 background imageLoading...
Page #57 background image
Output Configuration
Output Configuration
This section describes output channel configuration. Many commands associated with output configuration start with
SOURce1: or SOURce2: to indicate a certain channel. If omitted, the default is channel 1. For example, VOLT 2.5 sets
the output on channel 1 to 2.5 V, and SOUR2:VOLT 2.5 does the same for channel 2.
The instrument's display includes a "tab" for each channel that summarizes various aspects of each channel's output
configuration:
On a two-channel instrument, the tab for channel 1 will be yellow, and the tab for channel 2 will be green.
Output Function
The instrument includes eight standard waveforms: sine, square, ramp, pulse, triangle, noise, PRBS (pseudo-random
binary sequence), and DC. There are also nine built-in arbitrary waveforms, and you can create custom waveforms
with the embedded waveform editor.
The table below shows which functions are allowed (•) with modulation, sweep, and burst. Selecting a function that is
not allowed with a modulation or mode disables the modulation or mode.
Carrier AM FM PM FSK BPSK PWM Sum Burst Sweep
Sine and Square • • • • • • • •
Pulse • • • • • • • • •
Triangle and Ramp • • • • • • • •
Gaussian Noise • • •
a
PRBS • • • • •
Arbitrary Waveform • • •
b
•
b
• • •
Sequence • •
(a) Gated burst only
(b) Applies to sample clock, not whole waveform
l Frequency Limitations: Changing functions may change the frequency to meet the new function's frequency lim-
its.
l Amplitude Limitations: When the output units are Vrms or dBm, changing functions may lower the amplitude to
the maximum for the new function due to variation in waveform shapes. For example, a 5 Vrms square wave (into
50 Ω) changed to a sine will decrease to 3.536 Vrms (sine’s upper limit).
l Amplitude and offset cannot combine to exceed the instrument’s capability. The one you set last may be changed
to stay within limits.
l You may protect a device under test (DUT) by specifying upper and lower output voltage limits.
l Front Panel:
56 Agilent 33500 Series Operating and Service Guide

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies Agilent 33500 Series and is the answer not in the manual?

Agilent Technologies Agilent 33500 Series Specifications

General IconGeneral
BrandAgilent Technologies
ModelAgilent 33500 Series
CategoryInverter
LanguageEnglish

Related product manuals