Aborting a Calibration in Progress
Aborting a Calibration in Progress
Sometimes it may be necessary to abort a calibration in progress. You can abort a calibration at any time by turning off
the power or by issuing a remote interface device clear message followed by *RST.
The instrument stores calibration constants at the end of each adjustment procedure. If you lose power, or otherwise
abort an adjustment in progress, you will only need to perform the interrupted adjustment procedure again.
If power is lost when the instrument is attempting to write new calibration constants to mem-
ory, you may lose all calibration constants for the function. Typically, upon re-applying power,
the instrument will report error "-313, Calibration Memory Lost".