EasyManua.ls Logo

Agilent Technologies Agilent 33500 Series

Agilent Technologies Agilent 33500 Series
429 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
FUNCtion Subsystem
instrument's current settings. If you have manually edited a segment or sequence file such that the instrument
settings have been removed, the instrument settings will not be changed when you execute FUNCtion:ARBitrary.
l When you store an arbitrary waveform segment or sequence (MMEMory:STORe:DATA[1|2]), the instrument's cur-
rent settings (voltage values, sample rate, filter type, and so on) are stored in the segment or sequence file. When
you play the file for the first time with FUNCtion:ARBitrary, these settings are loaded and override the instru-
ment's current settings. If you have manually edited a segment or sequence file such that the instrument settings
have been removed, the instrument settings will not be changed when you execute FUNCtion:ARBitrary.
[SOURce[1|2]:]FUNCtion:ARBitrary:ADVance {TRIGger|SRATe}
[SOURce[1|2]:]FUNCtion:ARBitrary:ADVance?
Specifies the method for advancing to the next arbitrary waveform data point for the specified channel.
Parameter Typical Return
{TRIGger|SRATe},defaultTRIG TRIGorSRAT
Set advance methord to trigger:
FUNC:ARB:ADV TRIG
l TRIGger causes instrument to advance to next data point with each trigger received and forces
TRIGger[1|2]:SOURce to EXTernal.
l SRATe causes instrument to advance to next data point at the sample rate set by FUNCtion:ARBitrary:SRATe.
[SOURce[1|2]:]FUNCtion:ARBitrary:FILTer {NORMal|STEP|OFF}
[SOURce[1|2]:]FUNCtion:ARBitrary:FILTer?
Specifies the filter setting for an arbitrary waveform.
Parameter Typical Return
{NORMal|STEP|OFF},defaultSTEP NORMal,STEP, orOFF
Set filter to NORMal:
FUNCtion:ARBitrary:FILTer NORM
l NORMal filters the data points with the filter that provides the flattest frequency response. This effectively
smoothes the signal, but sharp transitions will have pre-shoot and overshoot.
l STEP filters the data points in a way that effectively smoothes the signal while minimizing the pre-shoot and over-
shoot. However, this setting has a narrower bandwidth than the NORMal setting.
l OFF steps from point to point at the sample rate. Moves between data points are accomplished as quickly as pos-
sible with no smoothing. If the <mode> is set to OFF, the instrument uses a filter whose bandwidth limit restricts
the maximum sample rate for the arbitrary waveform to 62.5 MSa/s.
Agilent 33500 Series Operating and Service Guide 213

Table of Contents

Related product manuals