INITiate Subsystem
Parameter Typical Return
Change both channels to the wait-for-trigger state:
INIT:IMM:ALL
l The trigger system is armed by INITiate[:IMMediate]. Once the trigger count is satisfied, the trigger system
returns to idle state and ignores further triggers. The triggered function will be left in whatever state is achieved
with the count of triggers. Rearming the trigger system with another INITiate[:IMMediate] allows further triggers
to apply.
l Use ABORt to return instrument to idle.
l If the specified channel has INIT:CONT set ON, these commands have no effect on trigger system and error -213
will be generated.
238 Agilent 33500 Series Operating and Service Guide