2 Power Meter Operation
42 E4418B Power Meter User’s Guide
5 The current setting of the reference calibration factor is displayed under
the
softkey. To change this setting, press . The power meter
displays the reference calibration factor in a pop- up window. Modify this
reference calibration factor (see below) as desired.
• Use or to modify the digit on which the cursor is currently
positioned.
• Use or to move to other digits.
6 To confirm your choice, press .
7 Press to calibrate the power meter. During calibration, the wait symbol
is displayed. (The power meter automatically turns on the POWER REF
output.)
8 Press . The current setting of the calibration factor is
displayed under the softkey. To change this setting, press .
The power meter displays the calibration factor in a pop- up window. Modify
this calibration factor (see below) as desired.
• Use or to modify the digit on which the cursor is currently
positioned.
• Use or to move to other digits.
9 To confirm your choice, press .
10 Connect the power sensor to the signal to be measured.
11 The measurement result is now displayed.