Power Meter Operation 2
E4418B Power Meter User’s Guide 43
Example
To make a measurement on channel A with a power sensor which has a
reference calibration factor of 99.8% and a calibration factor of 97.8% at the
measurement frequency.
• Disconnect the power sensor from any source.
• Press .
• Press .
• Press . Use the , , , and keys to enter 99.8. Press
.
• Connect the power sensor to the POWER REF output.
• Press .
• Press , . Use the , , , and keys to
enter 97.8. Press .
• Connect the power sensor to the signal to be measured.
• The measurement result is now displayed.