Power Meter Operation 2
E4418B Power Meter User’s Guide 43
Example
To make a measurement on channel A with a power sensor which has a
reference calibration factor of 99.8% and a calibration factor of 97.8% at the
measurement frequency.
Disconnect the power sensor from any source.
Press .
Press .
Press . Use the , , , and keys to enter 99.8. Press
.
Connect the power sensor to the POWER REF output.
Press .
Press , . Use the , , , and keys to
enter 97.8. Press .
Connect the power sensor to the signal to be measured.
The measurement result is now displayed.
Zero
Ref CF
%
Cal
Cal Fac
%