2 Power Meter Operation
58 E4418B Power Meter User’s Guide
6 When you have completed editing the frequency dependent offset table,
press
to return to the “Offset Tbls” screen.
NOTE
You can only edit the table name, you cannot delete it.
Done
NOTE
Ensure that the frequency points you use cover the frequency range of the signals you
want to measure. If you measure a signal with a frequency outside the frequency range
defined in the frequency dependent offset table, the power meter uses the highest or
lowest frequency point in the frequency dependent offset table to calculate the offset.