110 Agilent 1200 Series Autosampler Reference Manual
4 Repairing the Autosampler
Simple Repairs
The procedures described in this section can be done with the autosampler in
place in the stack. You will do some of these procedures on a more frequent
basis.
Table 16 Simple Repair Procedures
Procedure Typical Frequency Time Required Notes
Exchanging the needle
assembly
When needle shows indication of
damage or blockage
15 minutes See “Needle Assembly”on page 111
Exchanging the seat
assembly
When the seat shows indication of
damage or blockage
10 minutes See “Needle-Seat Assembly” on
page 114
Exchanging the rotor
seal
After approximately 30000 to 40000
injections, or when the valve
performance shows indication of
leakage or wear
30 minutes See “Rotor Seal” on page 120
Exchanging the
metering seal
When autosampler reproducibility
indicates seal wear
30 minutes See“Metering Seal and Plunger” on
page 124
Exchanging the gripper
arm
When the gripper arm is defective 10 minutes See“Gripper Arm” on page 128

Table of Contents

Other manuals for Agilent Technologies G1314B VWD

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies G1314B VWD and is the answer not in the manual?

Agilent Technologies G1314B VWD Specifications

General IconGeneral
BrandAgilent Technologies
ModelG1314B VWD
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals