Agilent 1200 Series Autosampler Reference Manual 167
Repairing the Autosampler 4
SUD Board
Frequency When defective.
Tools required Pozidrive No. 1 screwdriver.
Parts required SUD board G1313-66503.
Preparations for this
procedure
Remove the top cover (page 134).
Remove the transport assembly (page 139).
Remove the sampling unit (page 141).
CAUTION
Take care not to damage the flex board when removing the SUD board.
1 Disconnect all connectors from the SUD board. 2 Remove the two fixing screws, and lift out the board.
Fixing screws