168 Agilent 1200 Series Autosampler Reference Manual
4 Repairing the Autosampler
3 Install the new board. Ensure the board is positioned
between the board guide.
4 Secure the board in place with the two screws.
5 Reconnect all the connectors to the new board.
On completion of this procedure:
Install the sampling unit (page 141).
Install the transport assembly (page 139).
Install the top cover (page 134).Verify the transport
assembly alignment (page 102).
Fixing screws
Leak sensor
Metering drive
Injection
Valve
Needle drive
Flex board