Agilent 1200 Series Autosampler Reference Manual 71
Troubleshooting and Test Functions 3
Valve to Bypass Failed
The injection valve failed to switch to the bypass position.
The switching of the injection valve is monitored by two microswitches on the
valve assembly. The switches detect the successful completion of the valve
movement. If the valve fails to reach the bypass position, or if the microswitch
does not close, the error message is generated.
Probable Causes
Defective injection valve.
Defective ASM board.
Suggested Actions
Exchange the injection valve.
Exchange the ASM board.