16 Agilent 1200 Series Autosampler Reference Manual
1 Installing the Autosampler
Environment
Your autosampler will work within specifications at ambient temperatures
and relative humidity as described in Table 1.
CAUTION
Do not store, ship or use your autosampler under conditions where temperature
fluctuations may cause condensation within the autosampler. Condensation will
damage the system electronics. If your autosampler was shipped in cold weather, leave
it in its box, and allow it to warm up slowly to room temperature to avoid condensation.
Table 1 Physical Specifications - Autosampler (G1329A / G2260A)
Type Specification Comments
Weight 14.2 kg (31.3 lbs)
Dimensions
(height × width × depth)
200 × 345× 435 mm
(8 × 13.5 × 17inches)
Line voltage 100 – 120 or 220 – 240 VAC,
± 10 % Wide-ranging capability
Line frequency 50 or 60 Hz, ± 5%
Power consumption (apparent power)
Power consumption (active power)
300 VA
200 W
Maximum
Maximum
Ambient operating temperature 4 – 55 °C (41 – 131 °F) see on page 17
Ambient non-operating temperature -40 to 70 °C (-4 to 158 °F)
Humidity < 95 %, at 25 – 40 °C (77 – 104 °F) Non-condensing
Operating Altitude Up to 2000 m (6500 ft)
Non-operating altitude Up to 4600 m (14950 ft) For storing the
autosampler
Safety standards: IEC, CSA, UL Installation Category II, Pollution Degree 2
For inddor use only
WARNING
Using the autosampler at environmental temperatures higher than 50°C (122 °F)
may cause the rear panel to become hot.