Agilent 1200 Series Autosampler Reference Manual 75
Troubleshooting and Test Functions 3
Missing Vial
No vial was found in the position defined in the method or sequence.
When the gripper arm picks a vial out of the sample tray, the processor
monitors the gripper motor encoder. If a vial is present, the closing of the
gripper fingers is limited by the vial. However, if no vial is present, the gripper
fingers close too far. This is sensed by the processor (encoder position),
causing the error message to be generated.
Probable Causes
No vial in the position defined in the method or sequence.
Incorrect gripper alignment.
Defective gripper assembly (defective gripper fingers or belt).
Defective transport assembly flex board.
Suggested Actions
Install the sample vial in the correct position, or edit the method or
sequence accordingly.
Align gripper.
Exchange the gripper assembly.
Exchange the transport assembly.

Table of Contents

Other manuals for Agilent Technologies G1314B VWD

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies G1314B VWD and is the answer not in the manual?

Agilent Technologies G1314B VWD Specifications

General IconGeneral
BrandAgilent Technologies
ModelG1314B VWD
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals