76 Agilent 1200 Series Autosampler Reference Manual
3 Troubleshooting and Test Functions
Initialization Failed
The autosampler failed to complete initialization correctly.
The autosampler initialization procedure moves the needle arm and transport
assembly to their home positions in a predefined sequence. During
initialization, the processor monitors the position sensors and motor encoders
to check for correct movement. If one or more of the movements is not
successful, or is not detected, the error message is generated.
Probable Causes
Mechanical obstruction.
Defective sampling unit flex board.
Defective transport assembly flex board.
Defective sampling unit motor.
Defective ASM board.
Suggested Actions
Ensure unobstructed movement of the transport assembly.
Exchange the defective sampling unit motor.
Exchange the transport assembly.
Exchange the ASM board.

Table of Contents

Other manuals for Agilent Technologies G1314B VWD

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies G1314B VWD and is the answer not in the manual?

Agilent Technologies G1314B VWD Specifications

General IconGeneral
BrandAgilent Technologies
ModelG1314B VWD
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals