Agilent 1200 Series Autosampler Reference Manual 69
Troubleshooting and Test Functions 3
Restart Without Cover
The autosampler was restarted with the top cover and foam open.
The sensor on the ASM board detects when the top foam is in place. If the
autosampler is restarted with the foam removed, the autosampler switches
OFF within 30 s, and the error message is generated.
Probable Causes
autosampler started with the top cover and foam removed.
Suggested Actions
Replace the top cover and foam.