Agilent 1200 Series Autosampler Reference Manual 141
Repairing the Autosampler 4
Sampling Unit
Frequency When accessing internal parts, or when defective.
Tools required 1/4 inch wrench (supplied in accessory kit).
Flat-head screwdriver.
Parts required Sampling unit G1329-60008 for G1329A
Sampling unit G2260-60008 for G2260A
The sampling units come without injection valve and analytical head
assembly (see “Sampling Unit Assembly” on page 178).
Preparations for this
procedure
Remove the front cover.
Remove the vial tray.
Remove the top cover (page 134).
Remove the Illumination assembly (page 138)
Remove the transport assembly (page 139).
1 Remove the pump and column thermostat capillaries
(ports 1 and 6) from the injection valve.
2 Pull out the waste tube from the holder in the leak tray.