Agilent 1200 Series Autosampler Reference Manual 169
Repairing the Autosampler 4
Power Supply
Frequency When defective
Tools required Pozidrive No. 1 screwdriver
Parts required Power supply 0950-2528.
Preparations for this
procedure
Remove the top cover (page 134).
Remove the transport assembly (page 139).
Remove the sampling unit (page 141).
Remove the ASM board (page 157).
Remove the fan (page 155).
CAUTION
The ASM board is sensitive to electrostatic discharge. Always use the ESD strap (see
“Using the ESD Strap” on page 107) when handling electronic boards.
1 Remove the bottom foam. 2 Remove the power switch light pipe from the coupler.
CellFrame
Light
pipe
Coupler

Table of Contents